Spektometry

Renomowanej firmy Labtex

Spektometry

Spektometry

LTS-UV1810 Single Beam Grating 1200 UV VIS Spectrophotometer

LTS-UV1810 Single Beam Grating 1200 UV VIS Spectrophotometer


LTS-UV1810 series of instruments is easy to perform by simple parameter setting, photometric analysis, quantitative analysis, kinetic analysis, spectral scanning, multi-wavelength measurement, and DNA/protein analysis.
★The instrument is equipped with a disc rotating automatic eight-link sample holder with good reliability and centered light spot (compared with the double-row automatic eight-link holder), thus the instrument has high photometric accuracy

 

Labtex Biotech China Co., Ltd.Labtex Biotech China Co., Ltd.Features

1. Adopt the latest ARM system, built-in Chinese and English operating system, can store transfer data through USB interface.

2. Powerful, the host can independently complete photometric measurement, quantitative weighing, spectral scanning, kinetics, DNA/protein testing, multi-wavelength testing and data printing, etc.

3. The instrument adopts 320×240 dot matrix high brightness 6-inch LCD display, directly displaying the scan spectrum with simple screen interface.

4. The unique design of the optical system, high-performance 1200 bars / mm grating and imported receivers to ensure that the instrument has excellent performance indicators.

5. Socket deuterium lamp and tungsten lamp design, free of optical debugging after changing the lamp.

6. 24-bit high-speed, high-precision A/D conversion, high precision instrument, fast response time.

7. Wide sample chamber, can accommodate 5-100mm cuvettes of various specifications.

8. It can be directly connected to printer to print the graph and experimental data.

9. The instrument adopts suspension optical system design and long optical path design, which greatly improves the stability and reliability of the instrument and the optical resolution of the instrument.

10. Built-in bilingual operating system in Chinese and English, convenient for you to use.

 

 

Parameters

Model

LTS-UV1810

LTS-UV1810S

Optical System

Single bean, grating 1200lines/mm

Wavelength Range

190 nm- 1100nm

Spectral Bandwidth

1nm

0.5/1/2/4/5 nm

Wavelength Accuracy

±0.1nm(656.1nm D2); ≤±0.3nm(full wavelength range)

Wavelength Repeatability

≤0.1nm

Photometric Accuracy

±0.002A(0-0.5A), ±0.004A(0.5-1A), ±0.2%T(0-100%)

Photometric Repeatability

±0.001A(0-0.5A), ±0.002A(0.5-1A), ≤0.15%T(0-100%)

Photometric Range

0-200%T, -4-4A

Stray light

≤0.03%T

Stability

±0.0005A(in 500nm)

Baseline Flatness

±0.0015A

Noise

0.05%T

Scanning Speed

Fast, Mid, Slow

Light Source

Deuterium &Tungsten -halogen lamp

Display

320*240 LCD

Photometric Mode

T,A,C,E

Detector

Imported Silicon Photodiode

Output

USB port & Parallel port (Printer)

Power

AC 220V/50Hz or AC 110V/6OHz

Dimension(mm)

590*475*250

N.W(kg)

20

Packing Size(mm)

770*630*340

G.W(kg)

26