Feature:
1)Full Automatic sterilization process,easy to operate.
2)With drying function,Suitable for drying dressings.
3) With over temperature and pressure protection device.
4)The door opening mechanism cannot be operated until the pressure in the chamber is reduced to 0.027MPa.
It can’t be start on if the door doesn’t close tightly.
5)The safety valve will be open automatically when the inner pressure over 0.24MPa,and the steam be exhausting
to the water tank.
6)When the machine runs out of water ,the machine’s power and water will be cut-off and alarm
7)The chamber of the sterilizer is made of stainless steel.
8)with printing function.
Parameter
Technical data | LTA-150HB | LTA-200HB | LTA-280HB |
Sterilizing chamber volume | 150L φ440×1000mm | 200L φ515×1000mm | 280L φ600×1000mm |
working pressure | 0.22 MPa |
working temperature | 134℃ |
Adjustment of temperature | 40℃ – 134℃ |
Time of sterilization | 0-99 min |
Time of drying | 0-99 min |
Heat average | ≤± 2℃ |
Power | 9KW/380V 50Hz | 9KW/380V 50Hz | 12KW/380V 50Hz |
Dimension | 1400×650×1600 | 1400×750×1700 | 1520×910×1900 |
Transport dimension | 1560×820×1820 | 1560×910×1880 | 1680×1080×2100 |
G.W/N.W | 430/340 KG | 436/350 KG | 570/462 KG |