Renomowanej firmy Labtex

Specifications:
| Model | LTS-AAS3624 | LTS-AAS3628 | |
| Optics System | Type | Single Beam | |
| Monochromator | Anaberration C-T type, focal length 350mm | ||
| Dispersing element | optical grating 1800 bars /mm, flicker wavelength 250nm | ||
| Spectral bandwidth (nm) | 0.1、0.2、0.4、0.7、1.4、2nm six grades automatic switchover | ||
| Wavelength range (nm) | 185-900nm | ||
| Wavelength indication error | ±0.3 nm | ±0.2 nm | |
| Wavelength repeatability (nm) | ≤0.1nm | ||
| Spectral bandwidth deviation | ±0.02 nm | ||
| Light point | 4 | 8 | |
| Element lamp type | Ordinary element lamp | ||
| Photometric Property | Reading mode | Transmittance, absorbance, concentration | |
| Photometric Range | 0-125%,-0.1-3.00A | ||
| Cu static baseline drift | ±0.004A/30min | ±0.003A/30min | |
| Cu dynamic baseline shift | ±0.006A/15min | ±0.005A/15min | |
| Background correction | Deuterium lamp background correction capability >= 30x (1Abs) | Deuterium lamp >= 40x (1Abs) | |
| Graphite Furnace Analysis | Characteristic concentration | Cd ≤0.9pg Cu≤20pg | Cd≤0.6pg Cu≤20pg |
| Detection Limit | Cd≤1.5pg Cu ≤25pg | Cd≤1.0pg Cu≤25pg | |
| Repeatability(Cd) | RSD≤4% | RSD≤3% | |
| Graphite furnace temperature range | Room temperature ~ 3000℃ | ||
| Large power temperature rise range | 1500℃~3000℃ | ||
| Slope and hold time | 1s~255s | 1s~255s | |
| Heating rate | Maximum heating rate 2000℃/ s | ||
| Atomization heating method | Light-controlled heating, time-controlled heating, general heating | ||
| Outside protective gas flow | 1 L/min | ||
| Tube protection gas flow | 4 adjustable levels (0, 50, 200, 250ml/min flow) | ||
| safety precautions | Cooling water flow, protective gas pressure, furnace body temperature, power supply temperature, graphite tube installation alarm | ||
| Data processing | Measurement mode | Graphite furnace process | |
| Concentration calculation method | Standard curve method (6 linear and nonlinear fitting methods), standard addition method, interpolation method | ||
| Number of repeated measurements | 1-30 times, calculate and give the mean, standard deviation and relative standard deviation of absorbance and concentration | ||
| Report printing | Parameter printing, data result printing | ||
| Others | Computer | circumscribed | |
| Overall dimensions | 830*650*560 | ||
| Weight | 110Kg | ||
| Power | Voltage 220V±22V, frequency 50Hz ±1Hz | ||
| Operating temperature | 10℃~30℃ | ||
| Operating Humidity | 40%~85% | ||
Features

1. Excellent total reflection optical system ensures excellent signal-to-noise ratio of the instrument.
2. Comprehensive and safe intelligent gas control module.
3. Excellent full-band deuterium lamp background correction technology and high-performance self-priming dual background
correction.
4. Fully automatic instrument control, the light source supports high-performance element lamps, and can be quickly completed with one click.
5. Advanced eight-position lamp holder with 360-degree rotating design.
6. Novelly designed atomizer and intelligent lifting.
7. Simple, professional, automatic and fully operational RG-WinAAS workstation software under MS Window system.




Pomożemy w wyborze odpowiedniego inkubatora i odpowiemy na wszelkie nurtujące pytania.